Effects of atrial tachycardia and dehydration in an elderly patient with a leadless ventricular pacemaker implantation: a case report

  • Naoya Inoue
    Correspondence
    Corresponding author: Naoya Inoue Department of Cardiology, Chutoen General Medical Center 1-1 Shobugaike, Kakegawa, Shizuoka, Japan
    Affiliations
    Department of Cardiology, Chutoen General Medical Center, Kakegawa, Shizuoka, Japan

    Department of Cardiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan
    Search for articles by this author
  • Takashi Ogane
    Affiliations
    Department of Cardiology, Chutoen General Medical Center, Kakegawa, Shizuoka, Japan
    Search for articles by this author
  • Shuji Morikawa
    Affiliations
    Department of Cardiology, Chutoen General Medical Center, Kakegawa, Shizuoka, Japan

    Department of Cardiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan
    Search for articles by this author
Open AccessPublished:November 15, 2022DOI:https://doi.org/10.1016/j.hrcr.2022.11.006

      Keywords

      Key Teaching Points
      • ・When enhancing atrioventricular synchrony with the Micra AV, the possibility of dehydration and supraventricular tachycardia should be considered.
      • ・The A3/A4 signal is closely related to ventricular inflow velocity, and a threshold should be set when evaluating the E/A ratio.
      • ・To ensure adequate cardiac output, indications for the Micra AV should be carefully considered in frail and elderly patients, as they are more prone to dehydration.

      Introduction

      Recently, leadless pacemakers such as the Micra have become popular among patients who require pacemakers. Among Micra pacemakers, the Micra AV with VDD mode is used to maintain atrioventricular synchrony (AVS) in patients with an atrioventricular block (AVB). Complications associated with conventional transvenous pacemakers (TVPs)—such as device infection and subclavian vein stenosis—remain a concern, particularly in thin, elderly patients. The safety of Micra implantation in the elderly has already been reported.
      • Pagan E.
      • Gabriels J.
      • Khodak A.
      • et al.
      Safety of leadless pacemaker implantation in the very elderly.
      Additionally, although the Micra AV can provide AVS and can reduce pacemaker complications and maintain cardiac output,
      • Steinwender C.
      • Khelae S.K.
      • Garweg C.
      • et al.
      Atrioventricular synchronous pacing using a leadless ventricular pacemaker: results from the MARVEL 2 study.
      atrial mechanical sensing (AMS) using this device has various limitations.
      The feature of this pacemaker is that it does not have an atrial lead; thus, mechanical sensing of the blood flow due to atrial contraction (known as the A4 signal) provides appropriately timed ventricular pacing (VP). Pacemakers that do not maintain AVS are known to cause reduced cardiac output; therefore, considering that this is the right VP, it is suggested that implantation of the Micra AV may be disadvantageous. In this case report, a patient with AVB underwent Micra AV implantation and later developed atrial tachycardia (AT). Therefore, we set the A4 threshold to allow AMS; however, after discharge from the hospital, the patient developed activities of daily living (ADL) decline and worsening dehydration, and the Micra AV synchrony could not be maintained. This report highlights the need to consider whether the Micra AV is appropriate for use in elderly patients with AVB who have decreased activity.

      Case Report

      A 96-year-old woman presented with chronic heart failure due to aortic stenosis, hypertension, and chronic kidney disease due to nephrosclerosis. Approximately 6 months after first visiting the hospital, the patient was rushed to the emergency room due to loss of consciousness. An electrocardiogram (ECG) at the time of visit showed complete AVB, and blood tests on the same day revealed a uric acid (UA) level of 6.9 mg/dL, urea nitrogen (UN) level of 45.5 mg/dL, creatinine level of 2.1 mg/dL, and brain natriuretic protein (BNP) level of 319 pmol/L. Echocardiography (UCG) revealed: preserved left ventricular ejection fraction (LVEF: 55%); Emax, 1.03 m/s; Amax, 1.14 m/s (E/A: 0.91); deceleration time (DcT), 227 ms; E': 4.2 cm/s (Figure 1a). Physical examination revealed fluid retention, including pitting edema and jugular vein distention, and the patient was admitted to the hospital with heart failure due to AVB.
      Figure thumbnail gr1
      Figure 1Evaluation of E/A ratio by echocardiography (a) Before Micra AV implantation, (b) Appearance of atrial tachycardia and after dehydration, (c) After correction for dehydration by intravenous infusion, AV, atrioventricular; E/A, Emax/Amax
      The patient's advanced age and the need for pacemaker implantation as a therapeutic intervention for the underlying disease were thoroughly discussed with the patient's family. The patient strongly desired this treatment; however, due to her small build (height: 150.0 cm, weight: 40.8 kg, body mass index: 18.1 kg/m2) and the risk of infection due to advanced age, the Micra AV was selected instead of a TVP. The Micra AV was implanted in the right ventricular septum without any complications, and the pacemaker data was within normal parameters (threshold: 0.25 V/0.24 ms, sensing: 14.5 mV, and resistance: 960 Ω). During the Manual Atrial Mechanical (MAM) test after implantation, AT was observed (atrial rate: about 180 bpm). When the A3/A4 threshold was set to 4.5 m/s2 and 1.2 m/s2, respectively, good AMS and VP were obtained, and AVS was maintained.
      One week after discharge, the patient was rushed back to our hospital due to weakness caused by dehydration. Blood tests (UA level: 11.9 mg/dL, UN: 71.6 mg/dL, creatinine: 1.9 mg/dL, and BNP: 126 pmol/L) and second UCG were performed, and no significant change in left ventricular contractility was observed (LVEF: 61%). Changes were observed in the left ventricular inflow velocity waveforms: Emax, 0.83 m/s; Amax, 0.65 m/s (E/A: 1.27); DcT, 177 ms; and E', 3.7 cm/s (Figure 1b). ECG and MAM tests were performed; the ECG revealed persistent AT, and the interval between AMS and VP was irregular (Figure 2a). The MAM test indicated an undersensing of the A4 signal (Figure 3a); therefore, we resolved to treat the patient by administering supplemental fluids. Four days after supplementation, blood tests showed that dehydration had improved (UA level: 8.5 mg/dL, UN: 46.5 mg/dL, creatinine: 1.9 mg/dL, and BNP: 123 pmol/L); ECG, UCG, and MAM test results were reevaluated. The ECG results demonstrated that AVS was maintained (Figure 2b), and UCG showed improvement in the left ventricular inflow velocity wave and a decrease in the E/A ratio (Emax: 0.96 m/s, Amax: 0.91 m/s, and E/A 1.05; Figure 1c). The MAM test also showed that the A4 signal was sensed above the threshold (Figure 3b, c), and good AMS and VP were confirmed again (Figure 3b, d, e).
      Figure thumbnail gr2
      Figure 2Evaluation of atrioventricular synchrony by electrocardiography (a) After atrial tachycardia and dehydration, (b) After correction for dehydration by intravenous infusion, Arrow: A paced ventricular beat more than 300 ms from the ECG-confirmed P wave, Arrowhead: A regularly paced ventricular beat within 300 ms of the P wave, ECG, electrocardiography
      Figure thumbnail gr3
      Figure 3MAM test (a) After atrial tachycardia and dehydration (b) After correction for dehydration (c) Comparison of A4 amplitude after dehydration (left panel) and correction (right panel) (d) The transition of AM-VP and VP only rates (e) Histogram of AM-VP and VP only after dehydration correction, Arrow: undersensing and no AVS Arrowhead: A4 signal is good and AVS is obtained AVS, atrioventricular synchrony; AM-VP, atrial mechanical sensing-ventricular pacing; VP, ventricular pacing; MAM, manual atrial mechanical

      Discussion

      The Micra AV is a revolutionary pacemaker that enables mechanical sensing of atrial waves and appropriately timed VP by simply implanting the pacemaker in the ventricle. However, concerns about cardiac accidents associated with the Micra have been reported since its introduction, and maintaining perfect AVS with the Micra AV remains a challenge; the synchronization rate was reported to be approximately 70–80%, depending on the patient’s resting, exertion, and posture.
      • Chinitz L.
      • Ritter P.
      • Khelae S.K.
      • et al.
      Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: results from the micra atrioventricular feasibility studies.
      The importance of AV synchronization in pacemakers has been previously reported.

      Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 2019;140:e382-e482.

      A previous report demonstrated a statistically significant improvement in velocity-time integral compared with VVI and VDD using the Micra AV.
      • Steinwender C.
      • Khelae S.K.
      • Garweg C.
      • et al.
      Atrioventricular synchronous pacing using a leadless ventricular pacemaker: results from the MARVEL 2 study.
      Increasing stroke volume or cardiac output is particularly important in patients with heart failure; therefore, DDD-TVP is recommended when possible to achieve AVS. Nonetheless, the Micra AV is an important option for patients with background conditions, such as pacemaker infection or blood access problems due to hemodialysis.
      One prediction of AV synchrony in VDD correlates with the E/A ratio as reported by Garweg et al.
      • Garweg C.
      • Khelae S.K.
      • Steinwender C.
      • et al.
      Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: results from the MARVEL 2 study.
      In VDD, the A3 signal correlates with the E wave (the early diastolic wave of left ventricular inflow), and the A4 signal with the A wave (the atrial systolic wave).
      • Garweg C.
      • Khelae S.K.
      • Steinwender C.
      • et al.
      Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: results from the MARVEL 2 study.
      Previous reports have shown that a high AVS rate can be obtained when the E/A ratio is < 0.94 5; in this case, the preimplantation echocardiographic results predicted a high degree of AVS with VDD. In fact, the MAM test, which was performed immediately after implantation, demonstrated good AVS; however, subsequent dehydration may have led to the breakdown of interventricular synchrony. It is assumed that frequent atrial contraction due to AT reduced atrial contractility and the accompanying blood flow rate from the atria to the ventricles, which in turn reduced the single cardiac output. Additionally, the reduced venous perfusion due to dehydration may have further reduced blood flow into the right ventricle, resulting in mechanical sensing failure. As evidence for this mechanism, a comparison of the E wave, A wave, and E/A ratio before dehydration, after dehydration, and after correction for dehydration showed that the left ventricular inflow velocity wave and E/A ratio changed. Consistent with this, while the A4 threshold was lower for AT in the MAM test, the effect of dehydration still caused the A4 signal to be poorly sensed and AVS to fail.
      For such conditions, catheter ablation therapy for AT is one of the promising methods to improve AVS, but in this patient’s case, we decided to forego further invasive treatment due to her advanced age. It is also suggested that adequate hydration improves atrial pressure waveforms and AMS. Therefore, in this case, the AV ratio was set to follow the tachycardia at an AV ratio of 3:1 to maintain AVS as much as possible. However, the limitation is that the AT rate must not fluctuate, and if the cycle length (CL) fluctuates depending on whether focal AT or reentrant AT is used, there may be a risk of being unable to maintain AVS. Particularly, it was difficult to accurately evaluate the effect of dehydration on the atrial rate and CL; therefore, maintaining AVS in such conditions is a future challenge.
      As mentioned above, the Micra is a revolutionary device that improves the complications of conventional pacemakers. Still, the Micra AV should be carefully considered before its application, as its function may be affected by postimplantation events. Particularly, the Micra is currently recommended for less active rather than more active patients because of its current performance. Of course, battery life, device replacement, and additional devices, in the case of Micra, are required; however, from this perspective, the Micra may be suitable for the elderly at this time. However, it should be emphasized that the elderly, who are prone to ADL decline and consequent dehydration, may not benefit from Micra AV because dehydration may reduce AMS due to reduced blood flow into the ventricles. Conversely, the difficulty in tracking an AT may be encountered by both conventional TVP and Micra AV.

      Conclusions

      This study reports the case of an elderly patient with AVB who was predicted to have good AVS based on UCG; however, dehydration and AT events led to a loss of AVS. We strongly recommend that the decision to use the Micra AV be based on possible changes in the patient’s condition after implantation, not on the current activity assessment.

      Acknowledgments

      The authors thank physiologist Yuko Morishita for her assistance with echocardiography.
      The authors thank the team of clinical engineers, including Seigo Suzuki, Tomomi Matsuura, Takahisa Ishiyama, and Yuki Toda, who assisted with device checks.

      Funding and disclosures

      This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

      References

        • Pagan E.
        • Gabriels J.
        • Khodak A.
        • et al.
        Safety of leadless pacemaker implantation in the very elderly.
        Heart Rhythm. 2020; 17: 2023-2028
        • Steinwender C.
        • Khelae S.K.
        • Garweg C.
        • et al.
        Atrioventricular synchronous pacing using a leadless ventricular pacemaker: results from the MARVEL 2 study.
        JACC Clin Electrophysiol. 2020; 6: 94-106
        • Chinitz L.
        • Ritter P.
        • Khelae S.K.
        • et al.
        Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: results from the micra atrioventricular feasibility studies.
        Heart Rhythm. 2018; 15: 1363-1371
      1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 2019;140:e382-e482.