Atrial pacing induced oversensing in Subcutaneous Implantable Cardioverter Defibrillator (SICD)

Open AccessPublished:February 14, 2023DOI:https://doi.org/10.1016/j.hrcr.2023.01.016

      Keywords

      Introduction:

      The subcutaneous implantable cardioverter defibrillator (S-ICD) is an effective therapy for ventricular arrhythmias in patients who do not require either anti-bradycardia or anti-tachycardia pacing. The development of sinus or atrioventricular nodal conduction disease may present a challenge when pacemaker implantation is required.. This case report highlights the potential for device-device interaction.

      Case Report:

      An 18 year-old man with hypertrophic cardiomyopathy and a primary prevention S-ICD (SQ-RS 1010, Boston Scientific, Massachusetts, USA), developed first degree atrioventricular (AV) block (200-250ms) and right bundle branch block (RBBB) (QRS 134ms) with right axis deviation. The patient had not previously undergone a septal myectomy. A dual chamber pacemaker (PPM) was implanted (Epyra-6, Biotronik, Berlin, Germany). Prior to his S-ICD implant a transvenous ICD had previously been extracted after becoming infected after 5 years.
      The PPM was programmed DDD with a lower rate (LRL) of 60 bpm and an upper tracking rate (UTR) of 150 bpm. The S-ICD conditional and shock zones were set to 250 bpm in the secondary vector. Primary and alternate vectors were not viable options due to physiological oversensing of the patient's intrinsic sinus rhythm with RBBB. Prior to pacemaker implantation there was appropriate sensing by the S-ICD of the intrinsic R wave (figure 1).
      Figure thumbnail gr1
      Figure 1Three panels are presented; Panel A – a captured S-ICD electrogram showing appropriate sensing prior to implantation of concurrent pacemaker in the Secondary Configuration. S: Sense. Panel B- shows an anteroposterior chest radiograph. The S-ICD generator is located in the left mid-axillary line, with the S-ICD electrode tunneled from the pulse generator to the left parasternal region. Dual chamber pacemaker with IS-1, IS-1 header in the left pectoral region. Bipolar right atrial and right ventricular pacing leads. Panel C- a suggested optimal position of the S-ICD electrode straight and within 1 centimeter of the sternum and the S-ICD generator encompassing the ventricular mass.
      Two months post PPM insertion an untreated episode was detected by the S-ICD displaying intermittent triple counting of the P, R and T-wave (fig 2). The paced P-wave amplitude was similar to that of the intrinsic R-wave (fig. 2, red box) resulting in sensing of both components. A fortuitous ventricular ectopic (fig. 2, green box) reset the sensing profile as the R wave amplitude which was significantly greater and momentarily avoided further oversensing and inappropriate therapy.
      Figure thumbnail gr2
      Figure 2Electrogram’s of the untreated SICD Episode. Two panels are presented; Panel A - the recorded trace during the tachycardia episode above, Panel B - exploded sections of the trace. Panel A shows a continuous trace of the episode starting on the left of the top strip continuing to the strip below. The trace shows the patient in an atrial paced rhythm with intrinsic atrioventricular conduction. The markers below the trace show the device interpretation. The red box (exploded in panel B) shows a section of the trace in which the amplitude of the P, R and T-wave are of similar amplitude, the markers showing the device interpreting and detecting each component of the trace and starting the device counter. This continues (red arrow) until tachycardia detection is reached (conditional and shock zone set to 250 bpm) and the device charges (C). A ventricular ectopic (S) (green box exploded in Panel B) resets the auto gain control algorithm of the device, changing its sensitivity and improving its accuracy. S: Sense, T: Tachycardia detection, C: Charge start, N: Noise.
      Simultaneous interrogation of the PPM and S-ICD was performed. Whist conducting the atrial threshold test in AAI mode, the S-ICD P-wave oversensing was replicated (fig. 3, panel A). The PPM was re-programmed with a reduced LRL of 40 bpm and atrial auto-capture turned off to allow intrinsic P wave sensing. No further untreated episodes due to oversensing were seen. When the S-ICD generator reached recommended replacement time both systems were extracted and the patient was upgraded to a cardiac resynchronisation therapy defibrillator (CRT-D) in light of poor right ventricular (RV) function and severe heart failure and the risk of requiring increased/permanent RV pacing due to the prolonged AV delays.
      Figure thumbnail gr3
      Figure 3Captured S-ICD Electrograms during pacing. Three panels are presented; Panel A- captured trace when conducting atrial threshold in AAI in the case presented. The following two panels are from an additional example patient. Panel B- the recorded trace during atrial pacing and ventricular pacing at 60 bpm in the secondary configuration with intermittent oversensing of the terminal portion of the QRS and atrial pacing stimulus. Panel C- the recorded trace during atrial pacing and ventricular pacing at 100 bpm with intermittent P wave oversensing. S: Sense, Dot: Discarded beat

      Discussion:

      We believe this is the first reported case to highlight oversensing due to atrial pacing in a patient with an S-ICD and concurrent dual chamber pacemaker. While there was appropriate sensing by the S-ICD prior to the pacemaker implant, atrial pacing resulted in oversensing in the secondary configuration, which was the only configuration in which inappropriate sensing of myopotentials did not occur pre-pacemaker implant.
      Inappropriate sensing by the S-ICD occurred due to the similarity in amplitude of the P, R and T wave during atrial pacing with intrinsic atrioventricular conduction. As in the case of transvenous ICD’s the S-ICD utilises auto gain control whereby the average amplitude of the last two sensed signals is taken and the decay to sensing floor begins at 75% of this calculated amplitude. As the device consistently sensed the P wave, R wave and T wave, the interval between the sensed intervals reduced. Consequently the average of the last 4 R-R intervals, in which heart rate is calculated is greater than 167 bpm, therefore the sensing profile alters as it falls within the mid-rate zone. This resulted in a more aggressive decay to the sensing floor and shortened refractory period which caused continued oversensing. The alteration in sensing profiles with rates >167bpm or within the conditional or shock zone is to optimise detection of ventricular fibrillation. Consequently, a “tachycardia episode” was detected and the device began to charge. The ventricular ectopic triggered the auto gain control algorithm to reset due to its significantly greater signal amplitude, therefore the decay to sensing floor was less aggressive and no longer oversensing small amplitude signals and prevented potential inappropriate therapy.
      Upon review of the chest x-ray (fig. 1, panel B) the S-ICD generator and S-ICD electrode position were found to be sub-optimal and not in a position that would be accepted nowadays. However, for the first generation S-ICD devices (model 1010) implantation x-ray screening was not a recommendation made by the manufacturer, the position was based on satisfactory defibrillation threshold testing (DFT) and in-range shock impedance alone. It is unlikely that repositioning the S-ICD electrode or S-ICD generator to an optimal position as suggested in fig.1, panel C would resolve the over sensing issue. With this reasoning the decision was taken not to undertake lead/generator re-positioning.
      There are limited reports of pacemaker and S-ICD interaction in the literature. Kossidas et al, 2017 [
      • Kossidas K.
      • Kalman R.
      • Follis W.P.
      • Cooper J.M.
      Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system.
      ] described a case of an S-ICD and a dual chamber pacemaker with epicardial leads. Following implant device interaction was assessed with atrial sensing and ventricular pacing at 90 bpm with different AV delays and pacing outputs. They found with pacing outputs of 7.5V the unipolar pacing spikes were sensed and lead to double counting. Huang et al, 2016 [
      • Huang J.
      • Patton K.K.
      • Prutkin J.M.
      Concomitant Use of the Subcutaneous Implantable Cardioverter Defibrillator and a Permanent Pacemaker.
      ] discussed the evaluation of sensed and paced beats in different vectors intraoperatively and post implant. They suggested the PPM upper rate limit could be adjusted to less than or equal to 50% of the S-ICD tachycardia zones to minimise the risk of inappropriate shocks.
      Important learning points have been identified in a patient with an S-ICD and concurrent PPM. Firstly, it is pertinent to consider upgrade to transvenous ICD as it is not feasible to screen a patient pre-operatively with paced and sensed morphologies. When screening patients for an S-ICD in patients with a pre-existing PPM, it is necessary to review atrial sensing, atrial pacing, ventricular sensing and ventricular pacing at maximum outputs in a bipolar configuration. If the pacing lead is unipolar, caution must be taken to ensure double counting of the pacing spike and P/R wave is not present. Intraoperatively, during DFT pacing asynchronously at maximum output, is important so as to assess if the pacing artifact is sensed during arrhythmia to ensure in the VT/VF therapy is not delayed or aborted due to the inappropriate sensing. Finally, when programming the concurrent pacemaker, turning off the polarity switch and auto threshold testing algorithms can help limit oversensing by the S-ICD.
      In our case, the older model S-ICD did not have the SMART Pass™ algorithm available. SMART Pass™ is an additional high-pass digital filter that attenuates lower frequency signals whilst preserving higher frequency signals greater than 10Hz, designed to selectively filter out T waves and myopotentials. The same mechanism of oversensing was replicated in a different patient, however was not negated with the newer S-ICD model (Emblem A219) in which Smart Pass™ is a feature and was active (fig. 3, panel B&C).

      Conclusion:

      In patients who require an S-ICD and a concurrent pacemaker, the importance of screening all possible paced morphologies from both the atrial and ventricular chambers is required. Detailed knowledge of both device algorithms is essential to troubleshoot possible device interactions, especially as oversensing and other cross talk can only be evaluated after device implantation.

      Acknowledgements:

      Matthew Chapman and Peter Sawires

      References:

        • Kossidas K.
        • Kalman R.
        • Follis W.P.
        • Cooper J.M.
        Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system.
        HeartRhythm Case Rep. 2017 Sep1; 3 (579-58)
        • Huang J.
        • Patton K.K.
        • Prutkin J.M.
        Concomitant Use of the Subcutaneous Implantable Cardioverter Defibrillator and a Permanent Pacemaker.
        Pacing Clin Electrophysiology. 2016 Nov; 39: 1240-1245

      Linked Article